Anaesthesia System KAM-A102 is equipped with color LCD screen for clear display of trends, dosage rate and other setting parameters. Designed to suit adult as well as pediatric patients. System complies with closed, semi-closed, semi-open or assisted ventilation. Integrated system software allows real-time monitoring for loops, Paw, BPM, VT and lung compliance. Comprehensive audio-visual alarm indication for critical events increases patient safety.
Drive | Electronically controlled, pneumatic driven |
Tidal Volume | 20 ml – 1500 ml |
Configuration | |
Gas Configuration | O2, N2O |
Flow Meter | Cascade, 0 L/min -10 L/min |
LCD Display | 7 inch LCD |
Graphic Display | Waveforms of P-T, F-T |
Vaporizer | Isoflurane / Enflurane / Sevoflurane / Halothane |
Fresh Gas Compensation | |
I:E Range | 4:1-1:8 |
Pressure Range | 5 mbar - 80 mbar |
P trigger Range | -20 cmH2O to 20 cmH2O |
Monitoring | |
Modes of Ventilation | |
Respiration range | 1 bpm - 99 bpm (1 bpm - 40 bpm under SIMV) |
Concentration Range | |
Calibration | Automatic |
Pipeline Input | 280 kPa - 600 kPa |
O<sub>2</sub> Flush | |
Minute Volume | > 18 L/min |
PEEP | Off 3 - 20 cm H2O |
Run-time | > 30 minutes up to 2 hours |
Waveforms | |
Hypoxic Safety System | N2O cut-off valve / O2 > 25% |
Alarm | |
Pressure Safety Valve | < 12.5 k Pa |
Volume of CO<sub>2</sub> absorber | 1.5 L |
Output Pressure | 0.4 ± 0.1 mPa |
Power Consumption | |
Power Supply | AC 220V ± 10%, 50Hz ± 2% |
Battery | Lead-acid battery 12V, 4Ah |
Current | 500 mA (Max) |
Working Condition | |
Dimensions | 700 x 850 x 1400 mm |
Weight | 65 Kgs |
Used in Surgical or Gynaecological departments to administer appropriate anaesthesia dosage at a specified rate to ensure effective sedation during surgeries and other invasive diagnostic or interventional procedures.